We investigated with Monte Carlo simulations, using the EGSNrcMP code, the capabilities of a small animal PET scanner based on four stacks of double-sided silicon strip detectors. Each stack consists of 40 silicon detectors with dimension of 60×60×1 mm^3 and 128 orthogonal strips on each side. Two coordinates of the interaction are given by the strips, whereas the third coordinate is given by the detector number in the stack. The stacks are arranged to form a box of 5×5×6 cm^3 with minor sides opened; the box represents the minimal FOV of the scanner. The performance parameters of the SiliPET scanner have been estimated giving a (positron range limited) spatial resolution of 0.52 mm FWHM, and an absolute sensitivity of 5.1% at the center of system. Preliminary results of a proof of principle measurement done with the MEGA advanced Compton imager using a ≈1 mm diameter 22Na source, showed a focal ray tracing FWHM of 1 mm.
SiliPET: An ultra-high resolution design of a small animal PET scanner based on stacks of double-sided silicon strip detector
DI DOMENICO, Giovanni;ZAVATTINI, Guido;CESCA, Nicola;AURICCHIO, Natalia;
2007
Abstract
We investigated with Monte Carlo simulations, using the EGSNrcMP code, the capabilities of a small animal PET scanner based on four stacks of double-sided silicon strip detectors. Each stack consists of 40 silicon detectors with dimension of 60×60×1 mm^3 and 128 orthogonal strips on each side. Two coordinates of the interaction are given by the strips, whereas the third coordinate is given by the detector number in the stack. The stacks are arranged to form a box of 5×5×6 cm^3 with minor sides opened; the box represents the minimal FOV of the scanner. The performance parameters of the SiliPET scanner have been estimated giving a (positron range limited) spatial resolution of 0.52 mm FWHM, and an absolute sensitivity of 5.1% at the center of system. Preliminary results of a proof of principle measurement done with the MEGA advanced Compton imager using a ≈1 mm diameter 22Na source, showed a focal ray tracing FWHM of 1 mm.I documenti in SFERA sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.