Microelectrode arrays are used for recording and stimulation in neurosciences both in vitro and in vivo. The electrodeposition of conductive polymers, such as poly(3,4-ethylene dioxythiophene) (PEDOT), is widely adopted to improve both the in vivo recording and the charge injection limit of metallic microelectrodes. The workhorse of conductive polymers in the neurosciences is PEDOT:PSS, where PSS represents polystyrene-sulfonate. In this paper, the counterion is the fluorinated polymer Nafion, so the composite PEDOT:Nafion is deposited onto a flexible neural microelectrode array. PEDOT:Nafion coated electrodes exhibit comparable in vivo recording capability to the reference PEDOT:PSS, providing a large signal-to-noise ratio in a murine animal model. Importantly, PEDOT:Nafion exhibits a minimized polarization during electrical stimulation, thereby resulting in an improved charge injection limit equal to 4.4 mC cm(-2), almost 80% larger than the 2.5 mC cm(-2) that is observed for PEDOT:PSS.
Electrodeposited PEDOT:Nafion Composite for Neural Recording and Stimulation
Carli S.
Primo
Membro del Collaboration Group
;Bianchi M.Secondo
Membro del Collaboration Group
;Zucchini E.;Fadiga L.Penultimo
;Biscarini F.
2019
Abstract
Microelectrode arrays are used for recording and stimulation in neurosciences both in vitro and in vivo. The electrodeposition of conductive polymers, such as poly(3,4-ethylene dioxythiophene) (PEDOT), is widely adopted to improve both the in vivo recording and the charge injection limit of metallic microelectrodes. The workhorse of conductive polymers in the neurosciences is PEDOT:PSS, where PSS represents polystyrene-sulfonate. In this paper, the counterion is the fluorinated polymer Nafion, so the composite PEDOT:Nafion is deposited onto a flexible neural microelectrode array. PEDOT:Nafion coated electrodes exhibit comparable in vivo recording capability to the reference PEDOT:PSS, providing a large signal-to-noise ratio in a murine animal model. Importantly, PEDOT:Nafion exhibits a minimized polarization during electrical stimulation, thereby resulting in an improved charge injection limit equal to 4.4 mC cm(-2), almost 80% larger than the 2.5 mC cm(-2) that is observed for PEDOT:PSS.File | Dimensione | Formato | |
---|---|---|---|
carli2019.pdf
solo gestori archivio
Descrizione: editoriale
Tipologia:
Full text (versione editoriale)
Licenza:
NON PUBBLICO - Accesso privato/ristretto
Dimensione
1.59 MB
Formato
Adobe PDF
|
1.59 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in SFERA sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.